Most device teams are budgeting a second compliance programme for the AI in their product.
The text does not ask for one.
For AI inside a regulated device, the AI Act slots into the file you already keep for your notified body. The real work is mapping new requirements, clause by clause, into documentation, a QMS and a post-market plan that already exist.
If you already work with a notified body, this part of the Act was written for you.
Which AI in a medical device does the AI Act's product-safety route actually catch?
The product-safety route is Article 6(1). It makes an AI system high-risk only where two conditions are both met.
First, under Article 6(1)(a), "the AI system is intended to be used as a safety component of a product, or the AI system is itself a product, covered by the Union harmonisation legislation listed in Annex I."
Second, under Article 6(1)(b), the product "is required to undergo a third-party conformity assessment, with a view to the placing on the market or the putting into service of that product."
Both. Not either.
The Act names device law directly: Article 73(10) refers to AI systems "which are safety components of devices, or are themselves devices, covered by Regulations (EU) 2017/745 and (EU) 2017/746", the MDR and IVDR.
The second condition is where classification starts to matter. Under MDR Article 52(3) and (4), class III and class IIb devices go through conformity assessment procedures in Annex IX (or alternatives) that involve a notified body. Class IIa does too, under Article 52(6).
Class I is different. MDR Article 52(7) has class I manufacturers "declare the conformity of their products by issuing the EU declaration of conformity" after drawing up the technical documentation. A plain class I device is therefore likely outside Article 6(1)(b). That is an inference, not a ruling, and it has exceptions: the same paragraph sends class I devices "placed on the market in sterile condition, have a measuring function or are reusable surgical instruments" into notified-body procedures, with that involvement limited to specific aspects.
So not all device AI is high-risk. Test each product against both conditions rather than assume.
Nothing below turns on the application date. The file work is the same whenever the Article 6(1) obligations apply to your product.
If the AI ships inside your device, are you the AI provider?
Usually, yes.
Article 25(3) says that for high-risk AI systems that are safety components of products covered by Section A of Annex I, "the product manufacturer shall be considered to be the provider of the high-risk AI system." It applies where the AI is "placed on the market together with the product under the name or trademark of the product manufacturer", or put into service under that name or trademark after the product is on the market.
The provider carries Article 16: Section 2 compliance, an Article 17 QMS, and "the relevant conformity assessment procedure as referred to in Article 43."
If you licensed the model from a third party, Article 25(4) gives you a contractual lever. The provider and the supplier of "an AI system, tools, services, components, or processes" used in the high-risk system "shall, by written agreement, specify the necessary information, capabilities, technical access and other assistance" needed for the provider to comply.
If your model contract does not deliver what your notified body will ask for, that is the first gap.
Does the AI Act require a second technical file, or one?
One.
Article 11(2) is explicit. Where a high-risk AI system related to a product covered by Section A of Annex I is placed on the market or put into service, "a single set of technical documentation shall be drawn up containing all the information set out in paragraph 1, as well as the information required under those legal acts."
Article 8(2) sets the principle behind it. To "ensure consistency, avoid duplication and minimise additional burdens," providers "shall have a choice of integrating, as appropriate, the necessary testing and reporting processes, information and documentation they provide with regard to their product into documentation and procedures that already exist" under the sectoral legislation.
In practice, the single file is your MDR or IVDR technical documentation plus the AI Act content, which under Article 11(1) "shall contain, at a minimum, the elements set out in Annex IV."
The work is a crosswalk. Every AI Act Section 2 requirement needs a home inside an existing file section, or a new section where none exists.
Will your MDR or IVDR notified body assess the AI as well?
That is the design. Article 43(3) says that for high-risk AI systems covered by Section A of Annex I, "the provider shall follow the relevant conformity assessment procedure as required under those legal acts." The Section 2 requirements "shall apply to those high-risk AI systems and shall be part of that assessment."
Some AI Act procedure rides along. Article 43(3) adds that "Points 4.3., 4.4., 4.5. and the fifth paragraph of point 4.6 of Annex VII shall also apply."
Whether your current notified body can do the AI part depends on its own notification. Article 43(3) entitles notified bodies notified under the sectoral acts "to control the conformity of the high-risk AI systems with the requirements set out in Section 2, provided that the compliance of those notified bodies with requirements laid down in Article 31(4), (5), (10) and (11) has been assessed in the context of the notification procedure under those legal acts."
That "provided that" is the question to put to your notified body now.
One narrow rule: a sectoral opt-out from third-party assessment via harmonised standards is usable only if standards or common specifications also cover "all requirements set out in Section 2."
Can your existing QMS and post-market surveillance plan carry the AI Act duties?
Yes, with conditions on each.
On the QMS, Article 17(3) says providers already subject to QMS obligations under sectoral Union law "may include the aspects listed in paragraph 1 as part of the quality management systems pursuant to that law." The Article 17(1) aspects still have to be identifiable inside it, from "procedures for the management of modifications" to data management and "procedures related to the reporting of a serious incident in accordance with Article 73." Training-data acquisition, labelling and retention are what device QMS procedures most often lack.
On post-market monitoring, Article 72(4) applies where "a post-market monitoring system and plan are already established" under the sectoral legislation. Providers "shall have a choice of integrating, as appropriate, the necessary elements described in paragraphs 1, 2 and 3 using the template referred in paragraph 3 into systems and plans already existing under that legislation, provided that it achieves an equivalent level of protection."
The plan must still "actively and systematically collect, document and analyse relevant data" on AI performance (Article 72(2)). A plan built around complaints and vigilance may not yet collect model performance data. That is the gap.
When does an AI malfunction in a device become reportable under the AI Act, and to whom?
Less often than device teams fear.
The AI Act defines a serious incident in Article 3(49) as an incident or malfunction that leads to death or serious harm to health, serious disruption of critical infrastructure, "the infringement of obligations under Union law intended to protect fundamental rights", or serious harm to property or the environment.
For device AI, Article 73(10) narrows that. Notification "shall be limited to those referred to in Article 3, point (49)(c)", the fundamental-rights limb, "and shall be made to the national competent authority chosen for that purpose by the Member States where the incident occurred."
The inference, not stated in terms, is that health-harm events stay on your MDR or IVDR vigilance channel and the AI Act adds a report only for the fundamental-rights category. Build triage on that split and check it with counsel.
Market surveillance follows the same logic. Article 74(3) makes the market surveillance authority for these systems "the authority responsible for market surveillance activities designated under those legal acts", with a Member State option to designate another authority if coordination is ensured. Article 74(4) disapplies the AI Act procedures in Articles 79 to 83 "where such legal acts already provide for procedures ensuring an equivalent level of protection and having the same objective."
Same regulators. Same channels. One extra triage question.
What happens when a learning model changes after certification?
This is where device teams need the most care.
Article 43(4) says high-risk AI systems that have been through conformity assessment "shall undergo a new conformity assessment procedure in the event of a substantial modification." Article 3(23) defines that as a change "not foreseen or planned in the initial conformity assessment" which affects compliance with the Section 2 requirements or modifies the intended purpose.
The exit is planning. For systems "that continue to learn after being placed on the market or put into service," changes "pre-determined by the provider at the moment of the initial conformity assessment and are part of the information contained in the technical documentation referred to in point 2(f) of Annex IV, shall not constitute a substantial modification."
Write the change envelope into the technical file before the initial assessment, and changes inside it are not substantial modifications. Leave it out, and every retrain is a question.
One more edge: under Article 25(1)(b), a third party that substantially modifies a high-risk system on the market becomes its provider.
Mapping it once
The Act points every AI duty at a structure you already run: the file, the notified body, the QMS, the post-market plan, the regulators.
It does not do the mapping. Each Section 2 requirement has to land on a named file section, owner and evidence.
That is the work Aigis is built to make traceable. One device control mapped to the MDR clause and the AI Act clause it satisfies, every obligation traced to a verbatim quote from the legal text, so the gap list is a query rather than a project. See how it works for device makers at agrc.ai/medical-devices, how exposure is modelled at agrc.ai/risk-management, and how the engine is built at agrc.ai/platform.
FAQ: EU AI Act and AI-enabled medical devices
Is every AI-enabled medical device high-risk under the AI Act? No. Article 6(1) requires both that the AI is a safety component of, or itself, a product covered by Annex I legislation, and that the product requires third-party conformity assessment. Devices that self-declare conformity are likely outside 6(1)(b), subject to the MDR Article 52(7) exceptions.
Do I need a separate AI Act technical file? No. Article 11(2) requires a single set of technical documentation containing both the AI Act information and the information required under the sectoral legislation.
Will my MDR or IVDR notified body assess the AI? Article 43(3) makes the Section 2 requirements part of the sectoral notified body conformity assessment, provided the body's compliance with Article 31(4), (5), (10) and (11) was assessed in its notification. Ask your notified body directly.
Does retraining a model trigger a new conformity assessment? Only if it is a substantial modification. Under Article 43(4), changes pre-determined at the initial assessment and documented under Annex IV point 2(f) do not count as one.